A simple, rapid, cost-effective and sensitive high-performance liquid chromatography method with diode array detection was developed and validated for the quantification of letermovir, a compound approved for prophylaxis of cytomegalovirus infection and disease in adult recipients of an allogeneic hematopoietic stem cell transplant. Sorafenib was used as internal standard. Samples were pre-treated by liquid-liquid extraction with tert-butyl methylether. Separation was achieved on a XTerra® RP18 column (150 × 2.1 mm, 5 µm) at 30 °C using gradient elution with a mobile phase of 20 mM ammonium bicarbonate pH 7.9 (mobile phase A) and acetonitrile:20 mM ammonium bicarbonate (9:1 v/v) (mobile phase B). Samples were eluted at a flow rate of 0.3 mL/min throughout the 20-min run. UV wavelength mode was used, letermovir and sorafenib were monitored at 260 nm. The calibration curve was linear in a concentration range of 25-5000 ng/mL with correlation coefficients ≥ 0.99. Intra-day and inter-day accuracy expressed as relative error were -11.4-20% and -7.96-10.62%, respectively. Precision expressed as coefficient of variation was 1.44-3.15% (intra-day) and 1.17-1.93% (inter-day). The method was successfully applied for analysis of 128 letermovir levels demonstrating its usefulness for letermovir monitoring in routine clinical practice.
Keywords: Allogenic hematopoietic stem cell transplantation; Cytomegalovirus; Diode array detection (DAD); High-performance liquid chromatography (HPLC); Letermovir; Therapeutic drug monitoring.
Copyright © 2020 Elsevier B.V. All rights reserved.