Retrievable inferior vena cava (IVC) filters are self-expanding metallic devices implanted in the IVC to prevent migration of thrombi from the deep veins of the legs and pelvis to the lungs. The risk of complications from the filters increases with duration of implantation, but retrieval may be difficult due to intimal hyperplasia around the components of the filter. In this study, the potential for delivery of radiofrequency (RF) energy to the IVC wall via the filter was investigated. IVC filters were deployed in 4 swine while attached to a snare connected to a 480 kHz RF generator. Energy ranging from 0 to 48 kJ was applied via the filter followed by re-sheathing and withdrawal of the filter while connected to a force measurement device. Resheathing forces for the 0-energy cohort and pooled data from the 6-24 kJ cohorts were 4.50±0.70 N and 4.50±0.75 N, respectively. Petechial hemorrhages and variable non-occlusive thrombi were noted in some cohorts including the 0-energy cohort, consistent with delivery and acute retrieval of an IVC filter. Histologically, the extent of RF-induced injury was subtle at 6 kJ with focal areas of homogenized collagen while the 12 kJ cohort showed segmental tissue charring with coagulation necrosis which was more extensive for the 24 kJ cohort. The 48 kJ energy caused more extensive and non-target organ damage. The study demonstrated feasibility of delivery of RF to the IVC wall via a deployed filter, supporting further study of the ability of local RF heating of the IVC wall to inhibit the neointimal hyperplasia or as an aid in retrieval.
Keywords: IVC filter retrieval; animal model; inferior vena cava filter; radiofrequency ablation; retrieval force; swine.