Background: A comprehensive, stable, and efficient high-performance liquid chromatography-tandem mass spectrometry method was developed for rapidly analyzing 14 antidepressants and 13 antipsychotics in human plasma for routine clinical therapeutic drug monitoring.
Methods: Simple protein precipitation was used for the pretreatment of plasma samples; dynamic multiple reaction monitoring was used to avoid the loss of sensitivity caused by numerous ion transitions. In all, 80 ion transitions of 40 compounds were quantitatively determined in 6 minutes.
Results: The limit of detection for the 27 analytes was in the range of 0.1-30 ng/mL, and all calibration lines prepared using blank plasma were linear with a correlation coefficient of r2 ≥ 0.99. The method was accurate and precise with acceptable intraday and interday precisions (coefficients of variation, ≤20% for a lower limit of quantification and ≤15% for other quality control samples) and an accuracy of 85.51%-114.77%. This analysis method has been completely validated and successfully used in routine clinical therapeutic drug monitoring for more than 9963 samples [including 488 samples having drug concentrations above the laboratory alert level (supra-alert-level samples)] at Xiamen Xianyue Hospital.
Conclusions: This dynamic method is comprehensive (includes most antidepressants and antipsychotics listed in China), reliable (stably used for almost 2 years), and efficient (convenient sample processing and short run time) and provides a large amount of meaningful data for optimized pharmacotherapy. Our experimental data from the plasma concentrations of supra-alert-level samples could serve as a reference for the interpretation of the pharmacokinetics of patients with a high risk of toxicity or loss of tolerability.
Copyright © 2021 Wolters Kluwer Health, Inc. All rights reserved.