Improved performance and safety from Argus II retinal prosthesis post-approval study in France

Acta Ophthalmol. 2021 Nov;99(7):e1212-e1221. doi: 10.1111/aos.14728. Epub 2020 Dec 23.

Abstract

Purpose: To evaluate the post-approval long-term outcomes of the Argus II Retinal Prosthesis, with a specific focus on its functional visual benefit in patients' daily activities.

Methods: Eighteen patients with bare light perception due to end-stage retinitis pigmentosa were included in a French prospective, multicentre, single-arm study and followed for 2 years. Visual benefit in patients' daily activities was monitored through the use of the Functional Low-vision Observer Rated Assessment (FLORA), and the final score at 2 years was the primary effectiveness outcome. Standardized visual assessments were also performed. Device- or procedure-related adverse events were recorded.

Results: Seventeen subjects completed the study. Positive impacts of the Argus II system on functional vision and well-being were demonstrated for over 70% of subjects on the FLORA. Among the daily activities/tasks tested, finding doorways was one of the most statistically significantly improved tasks (p < 0.001), along with estimating the size of an obstacle (p < 0.001), visually locating a place setting on a dining table (p < 0.001) and visually locating people in a non-crowded setting (p < 0.001). Visual function was improved on most standardized tests. Only two device- or procedure-related serious adverse events were observed (one vitreous haemorrhage and one endophthalmitis, both resolved with treatment). No explantation was required.

Conclusion: This first report of a completed post-approval study of Argus II with a two-year follow-up demonstrates the safety and effectiveness of the Argus II System in a real-world cohort of patients and further highlights its real functional benefit in implanted patients' daily activities.

Keywords: Argus II; daily activities; quality of life; retinal implant; retinal prosthesis; retinitis pigmentosa.

Publication types

  • Multicenter Study

MeSH terms

  • Device Approval*
  • Female
  • Follow-Up Studies
  • France / epidemiology
  • Humans
  • Incidence
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Design
  • Prosthesis Implantation / methods*
  • Retina / physiopathology
  • Retina / surgery*
  • Retinitis Pigmentosa / epidemiology
  • Retinitis Pigmentosa / physiopathology
  • Retinitis Pigmentosa / surgery*
  • Time Factors
  • Visual Acuity*
  • Visual Prosthesis / standards*

Grants and funding