Objectives: Durability of sutureless aortic bioprosthetic valves remains a major issue. The aim of this study was to assess structural valve deterioration (SVD) and bioprosthetic valve failure (BVF) of the Perceval bioprosthesis using the new proposed standardized definitions.
Methods: All patients who underwent aortic valve replacement with sutureless Perceval S prostheses up to September 2016 were included. Clinical and echocardiographic follow-up was performed. New standardized definitions were used to assess the durability of sutureless bioprosthetic valves. From 2013 to 2016, 214 patients were included.
Results: The mean age and EuroSCORE II were 79 years and 2.74. Thirty-day mortality was 0.47%. The survival rate was 96.8%, 88.1% and 85.7% at 1, 3 and 4 years, respectively. The median echocardiographic follow-up was 3.28 years. The mean pressure gradient was 11.3 mmHg. No cases showed evidence of severe SVD, 17 patients had moderate SVD with a mean pressure gradient of 24 mmHg and 8 patients had definite late BVF. The incidence of moderate SVD and BVF at 4 years was 8.8% and 2.9%, respectively.
Conclusions: Mid-term follow-up to 6.3 years after aortic valve replacement with the Perceval bioprosthesis documents favourable haemodynamic and clinical outcomes and low rates of SVD and BVF.
Keywords: Extracorporeal circulation; Minimally invasive surgery; Valve disease.