Repeating the BinaxNOW antigen test for severe acute respiratory syndrome coronavirus 2 using 2 groups of readers within 30 minutes resulted in high concordance (98.9%) in 2110 encounters. Same-day repeat antigen testing did not significantly improve test sensitivity (77.2% to 81.4%) while specificity remained high (99.6%).
Keywords: BinaxNOW; COVID-19; SARS-CoV-2; antigen testing; diagnostic performance.