There is limited information about durability of large diameter porcine bioprostheses implanted for pulmonary valve replacement (PVR). We studied patients who underwent surgical PVR from 2002-2019 with a stented porcine bioprosthetic valve (BPV) with a labeled size ≥27 mm. The primary outcome was freedom from reintervention. During the study period, 203 patients underwent PVR using a porcine BPV ≥27 mm, 94% of whom received a Mosaic valve (Medtronic Inc., Minneapolis, MN). Twenty patients underwent reintervention from 3.4-12.0 years after PVR: 5 surgical and 15 transcatheter PVR procedures. The indication for reintervention was regurgitation in 13 patients, stenosis in 2, mixed disease in 4, and endocarditis in 1. Estimated freedom from reintervention was 97±1% at 5 years and 82±4% at 10 years, and freedom from prosthesis dysfunction (moderate or severe regurgitation and/or a maximum Doppler gradient ≥50 mm Hg) over time was 91±2% at 5 years and 74±4% at 10 years. Younger age and smaller true valve diameter were associated with shorter freedom from reintervention, but valve oversizing was not. The durability of large stented porcine bioprostheses in the pulmonary position is generally excellent, particularly in adolescents and adults, similar to various other types of BPV. In the current study, relative valve size was not associated with valve longevity, although the low event-rate in this population was a limiting factor.
Keywords: Pulmonary regurgitation; Pulmonary stenosis; Pulmonary valve replacement; Tetralogy of Fallot; Transcatheter valve.