Background and objectives: The brachial-ankle pulse wave velocity (baPWV) is one of the most widely used arterial stiffness variables for assessing vascular risk. The abiPWV is a device that calculates various PWVs and the blood pressure ankle-brachial index (ABI). The device can also determine baPWV. The aim of this study was to calculate the baPWV with abiPWV, validate it with a reference device (VaSera) and study its clinical usefulness.
Patients and methods: We studied 113 patients (mean age, 53 ± 12 years), 59 (52%) of whom were women, and 10 (8.8%) of whom had a previous cardiovascular event. The participants were classified according to cardiovascular risk factors (CRFs) into group I (none), group II (1 or 2 CRFs) and group III (3 or more CRFs). The patients with a previous cardiovascular event were included in group III. All participants had their baPWV measured with abiPWV and VaSera.
Results: The baPWV correlation between the 2 devices was r = 0.93 (p < .001), and the percentage error calculated with the Bland-Altman analysis was 4.5%. The baPWV measured with abiPWV (in m/s) was as follows: group I, 10.5 ± 1.6; group II, 13.8 ± 2.9 (p < .001 when compared with group I); and group III, 14.1 ± 2.7 (p < .001 when compared with group I). There were no differences between groups II and III. The results with VaSera were comparable to those of abiPWV.
Conclusions: Measuring baPWV with the abiPWV is safe and has a similar clinical utility to that of VaSera. Incorporating this function into the options of abiPWV makes it a complete device for assessing arterial stiffness.
Keywords: Arterial stiffness; Brachial-ankle pulse wave velocity; Cardiovascular risk factors; Factores de riesgo vascular; Rigidez arterial; VOPITB; Velocidad de onda de pulso brazo-tobillo; abiPWV.
Copyright © 2020. Published by Elsevier España, S.L.U.