Usability of smart infusion pumps: A heuristic evaluation

Appl Ergon. 2022 Jan:98:103584. doi: 10.1016/j.apergo.2021.103584. Epub 2021 Sep 22.

Abstract

Although smart infusion pumps were built to eliminate medication errors, new types of usability errors have arisen. The purposes of this study were to determine potential risks when using smart pumps during secondary medication administration and to identify opportunities for design improvements. We observed and analyzed nurses when they interacted with smart pumps and heuristically evaluated the smart pump to identify usability problems. Forty-three usability problems were identified with the smart pump. The usability problems have the potential to create high cognitive burden on nurses and to increase the likelihood of mistakes. We discuss design and process improvement recommendations for each major finding from this study.

Keywords: Human factors evaluation; Infusion pump safety; Medication errors; Roller clamp; Smart infusion pump; Smart pump usability.