Background: High adherence to remote monitoring (RM) in pacemaker (PM) patients improves outcomes; however, adherence remains suboptimal. Bluetooth low-energy (BLE) technology in newer-generation PMs enables communication directly with patient-owned smart devices using an app without a bedside console.
Objective: To evaluate the success rate of scheduled RM transmissions using the app compared to other RM methods.
Methods: The BlueSync Field Evaluation was a prospective, international cohort evaluation, measuring the success rate of scheduled RM transmissions using a BLE PM or cardiac resynchronization therapy PM coupled with the MyCareLink Heart app. App transmission success was compared to 3 historical "control" groups from the Medtronic de-identified CareLink database: (1) PM patients with manual communication using a wand with a bedside console (PM manual transmission), (2) PM patients with wireless automatic communication with the bedside console (PM wireless); (3) defibrillator patients with similar automatic communication (defibrillator wireless).
Results: Among 245 patients enrolled (age 64.8±15.6 years, 58.4% men), 953 transmissions were scheduled through 12 months, of which 902 (94.6%) were successfully completed. In comparison, transmission success rates were 56.3% for PM manual transmission patients, 77.0% for PM wireless patients, and 87.1% for defibrillator wireless patients. Transmission success with the app was superior across matched cohorts based on age, sex, and device type (single vs dual vs triple chamber).
Conclusion: The success rate of scheduled RM transmissions was higher among patients using the smart device app compared to patients using traditional RM using bedside consoles. This novel technology may improve patient engagement and adherence to RM.
Keywords: Digital health; Pacemaker; Remote monitoring; Smart devices; Telemedicine.
© 2021 Heart Rhythm Society. Published by Elsevier Inc.