Introduction: The Walrus balloon guide catheter (BGC) is a new generation of BGC, designed to eliminate conventional limitations during mechanical thrombectomy.
Objective: To report a multi-institutional experience using this BGC for proximal flow control (PFC) in the setting of carotid artery angioplasty/stenting (CAS) in elective (eCAS) and tandem strokes (tCAS).
Methods: Prospectively maintained databases at 8 North American centers were queried to identify patients with cervical carotid disease undergoing eCAS/tCAS with a Walrus BGC.
Results: 110 patients (median age 68, 64.6% male), 80 (72.7%) undergoing eCAS and 30 (27.3%) tCAS procedures, were included (median cervical carotid stenosis 90%; 46 (41.8%) with contralateral stenosis). Using a proximal flow-arrest technique in 95 (87.2%) and flow-reversal in 14 (12.8%) procedures, the Walrus was navigated into the common carotid artery successfully in all cases despite challenging arch anatomy (31, 28.2%), with preferred femoral access (103, 93.6%) and in monitored anesthesia care (90, 81.8%). Angioplasty and distal embolic protection devices (EPDs) were used in 91 (83.7%) and 58 (52.7%) procedures, respectively. tCAS led to a modified Thrombolysis in Cerebral Infarction 2b/3 in all cases. Periprocedural ischemic stroke (up to 30 days postoperatively) rate was 0.9% (n=1) and remote complications occurred in 2 (1.8%) cases. Last follow-up modified Rankin Scale score of 0-2 was seen in 95.3% of eCAS cohort, with no differences in complications in the eCAS subgroup between PFC only versus PFC and distal EPD (median follow-up 4.1 months).
Conclusion: Walrus BGC for proximal flow control is safe and effective during eCAS and tCAS. Procedural success was achieved in all cases, with favorable safety and functional outcomes on short-term follow-up.
Keywords: atherosclerosis; balloon; cervical; stroke.
© Author(s) (or their employer(s)) 2022. No commercial re-use. See rights and permissions. Published by BMJ.