Purpose: To evaluate the impact of magnetization transfer (MT) on brain tissue contrast in turbo-spin-echo (TSE) and EPI fluid-attenuated inversion recovery (FLAIR) images, and to optimize an MT-prepared EPI FLAIR pulse sequence to match the tissue contrast of a clinical reference TSE FLAIR protocol.
Methods: Five healthy volunteers underwent 3T brain MRI, including single slice TSE FLAIR, multi-slice TSE FLAIR, EPI FLAIR without MT-preparation, and MT-prepared EPI FLAIR with variations of the MT-preparation parameters, including number of preparation pulses, pulse amplitude, and resonance offset. Automated co-registration and gray matter (GM) versus white matter (WM) segmentation was performed using a T1-MPRAGE acquisition, and the GM versus WM signal intensity ratio (contrast ratio) was calculated for each FLAIR acquisition.
Results: Without MT preparation, EPI FLAIR showed poor tissue contrast (contrast ratio = 0.98), as did single slice TSE FLAIR. Multi-slice TSE FLAIR provided high tissue contrast (contrast ratio = 1.14). MT-prepared EPI FLAIR closely approximated the contrast of the multi-slice TSE FLAIR images for two combinations of the MT-preparation parameters (contrast ratio = 1.14). Optimized MT-prepared EPI FLAIR provided a 50% reduction in scan time compared to the reference TSE FLAIR acquisition.
Conclusion: Optimized MT-prepared EPI FLAIR provides comparable brain tissue contrast to the multi-slice TSE FLAIR images used in clinical practice.
Keywords: FLAIR; contrast; echo planar imaging; magnetization transfer.