Background and aims: Many knives have been developed to improve the efficacy and safety of endoscopic submucosal dissection (ESD). We aimed to evaluate the efficacy and safety of scissor-type knives for colorectal ESD compared with needle-type knives.
Methods: We performed a post-hoc propensity score-matched analysis in an 11-facility study between August 2013 and December 2018. A total of 2330 patients (2498 lesions) who underwent colorectal ESD were divided into needle-type (1923 patients, 2067 lesions) and scissor-type (407 patients, 431 lesions) knife groups. Short-term outcomes were compared between the 2 groups.
Results: Two-to-one propensity score-matched analysis identified 814 (709 patients) and 407 (386 patients) lesions in the needle- and scissor-type knife groups, respectively. The median resection speed was significantly faster in the needle-type group (18.3 mm2/min) than in the scissor-type group (13.2 mm2/min, P < .0001), whereas en-bloc and histologic complete resection rates were not significantly different between the needle- and scissor-type groups (96.8% [788/814] vs 98.3% [400/407], P = .1888 and 95.1% [774/814] vs 95.6% [389/407], P = .7763, respectively). The rate of lesions resected using a single knife was significantly higher in the scissor-type group (98.5% [401/407]) than in the needle-type group (43.9% [357/814], P < .0001). Rates of intraoperative perforation and delayed bleeding were significantly lower in the scissor-type group than in the needle-type group (.7% [3/407] vs 2.5% [20/814], P = .0431 for each).
Conclusions: Scissor-type knives are safer for colorectal ESD. However, they are associated with slower resection speeds compared with needle-type knives. (Clinical trial registration number: UMIN000016197.).
Copyright © 2022 American Society for Gastrointestinal Endoscopy. Published by Elsevier Inc. All rights reserved.