Objectives: The aim of this study was to review a single institution's experience with EXCOR Paediatric implantation.
Methods: Patients <15 years old who underwent EXCOR implantation as a bridge to transplantation between 2015 and 2021 were enrolled. Major adverse events included death, cerebrovascular event resulting in sequelae, major infection (sepsis or surgical site infection requiring open sternal irrigation or device removal) and device malfunction requiring surgical treatment.
Results: Overall median age and weight for all 20 children at implantation were 10.8 (interquartile range, 7.9-33.2) months and 6.3 (4.6-10.2) kg. Ten patients (50%) weighed <5 kg. Primary diagnoses were dilated cardiomyopathy in 13 patients, fulminant myocarditis in 3, restrictive cardiomyopathy in 2 and congenital heart disease in 2. Two patients required biventricular assist support. The median support time was 365 (241-636) days. Six patients (30%) were supported for >20 months. One patient died. Seven patients underwent heart transplant. Heart transplant has not been performed in the last 1.5 years. Five patients were weaned from EXCOR support after native myocardial recovery, including a patient with dilated cardiomyopathy who recovered after 24 months of EXCOR support. Major complication-free survival at 6, 12 and 18 months were 79.3%, 49.6% and 38.6%, respectively. Body weight <5 kg at implantation was a risk factor for decreased major complication-free survival.
Conclusions: Survival during EXCOR Paediatric support was good, but it prolonged the wait time for a heart transplant. The number of major complications increased over time and was not negligible, especially in small children.
© The Author(s) 2022. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery.