Background: Ventricular arrhythmias (VAs) are observed in 25%-50% of continuous-flow left ventricular assist device (CF-LVAD) recipients, but their role on mortality is debated.
Methods: Sixty-nine consecutive patients with a CF-LVAD were retrospectively analyzed. Study endpoints were death and occurrence of first episode of VAs post CF-LVAD implantation. Early VAs were defined as VAs in the first month after CF-LVAD implantation.
Results: During a median follow-up of 29.0 months, 19 patients (27.5%) died and 18 patients (26.1%) experienced VAs. Three patients experienced early VAs, and one of them died. Patients with cardiac resynchronization therapy (CRT-D) showed a trend toward more VAs (p = 0.076), compared to patients without CRT-D; no significant difference in mortality was found between patients with and without CRT-D (p = 0.63). Patients with biventricular (BiV) pacing ≥98% experienced more frequently VAs (p = 0.046), with no difference in mortality (p = 0.56), compared to patients experiencing BiV pacing <98%. There was no difference in mortality among patients with or without VAs after CF-LVAD [5 patients (27.8%) vs. 14 patients (27.5%), p = 0.18)], and patients with or without previous history of VAs (p = 0.95). Also, there was no difference in mortality among patients with a different timing of implant of implantable cardioverter-defibrillator (ICD), before and after CF-LVAD (p = 0.11).
Conclusions: VAs in CF-LVAD are a common clinical problem, but they do not impact mortality. Timing of ICD implantation does not have a significant impact on patients' survival. Patients with BiV pacing ≥98% experienced more frequently VAs.
Keywords: heart failure; left ventricular assist devices; mortality; ventricular arrhythmias.
© 2022 The Authors. Artificial Organs published by International Center for Artificial Organ and Transplantation (ICAOT) and Wiley Periodicals LLC.