Post-market evaluation is important to ensure the ongoing safety and effectiveness of cardiovascular implantable electronic device (CIED) leads. The Twenty-First Century Cures Act and subsequent Food and Drug Administrative (FDA) Guidance provide an opportunity to leverage real-world data sources for this purpose. The past 4 years have seen the development of EP PASSION: a multi-stakeholder, collaborative effort between the FDA, CIED manufacturers, Heart Rhythm Society, and academics. Using real-world data, EP PASSION enables longitudinal evaluation of the long-term safety of CIED leads, addressing limitations of current approaches to generate evidence that informs regulatory, clinical, and manufacturer decision-making. This state of the art article describes the impetus for and launch of EP PASSION, the lessons learned, its current state, the current analytic approach, and the strengths and limitations of leveraging extant data sources for post-market lead evaluation. We also compare EP PASSION to traditional post-approval studies and describe possible future directions.
Keywords: Adverse events; Cardiovascular implantable electronic devices; Leads; Post-market; Real-world evidence; Surveillance.
© 2022. This is a U.S. government work and not under copyright protection in the U.S.; foreign copyright protection may apply.