Precision of image-guided spinal stereotactic ablative radiotherapy and impact of positioning variables

Phys Imaging Radiat Oncol. 2022 May 5:22:73-76. doi: 10.1016/j.phro.2022.04.006. eCollection 2022 Apr.

Abstract

Background and purpose: Spinal stereotactic ablative body radiotherapy (SABR) requires high precision. We evaluate the intrafraction motion during cone-beam computed tomography (CBCT) guided SABR with different immobilization techniques.

Material and methods: Fifty-seven consecutive patients were treated for 62 spinal lesions with SABR with positioning corrected in six degrees of freedom. A surface monitoring system was used for patient set up and to ensure patient immobilization in 65% of patients. Intrafractional motion was defined as the difference between the last CBCT before the start of treatment and the first CT afterwards.

Results: For all 194 fractions, the mean intrafractional motion was 0.1 cm (0-1.1 cm) in vertical direction, 0.1 cm (0-1.1 cm) in longitudinal direction and 0.1 cm (0-0.5 cm) in lateral direction. A mean pitch of 0.6° (0-4.3°), a roll of 0.5° (0-3.4°) and a rotational motion of 0.4° (0-3.9°) was observed. 95.5% of the translational errors and 95.4% of the rotational errors were within safety range. There was a significantly higher rotational motion for patients with arms along the body (p = 0.01) and without the use of the body mask (p = 0.05). For cervical locations a higher rotational motion was seen, although not significant (p = 0.1). The acquisition of an extra CBCT was correlated with a higher rotational (pitch) motion (p = 0 < 0.01).

Conclusion: Very high precision in CBCT guided and surface-guided spinal SABR was observed in this cohort. The lowest intrafraction motion was seen in patients treated with arms above their head and a body mask. The use of IGRT with surface monitoring is an added value for patient monitoring leading to treatment interruption if necessary.

Keywords: Intrafraction motion; SABR; Spinal metastases.