Background: The mechanically-expandable transcatheter valve is no longer commercially available, yet clinical and echocardiographic surveillance is imperative for thousands of patients who received transcatheter aortic valve implantation (TAVI) with this platform.
Aims: We aimed to determine the incidence and mechanism of bioprosthetic valve dysfunction (BVD) following TAVI with mechanically-expandable valves.
Methods: From 2013 to 2020, all 234 patients who underwent TAVI with the LOTUS valve were included. BVD was categorised as (i) structural valve deterioration (SVD), (ii) non-structural valve dysfunction (NSVD), (iii) clinical valve thrombosis and (iv) endocarditis, according to the Valve Academic Research Consortium-3 criteria.
Results: The mean age was 79±7 years, 60% were male, and the mean Society of Thoracic Surgeons score was 4.2±2.9%. The technical success rate was 94% and the 30-day device success rate was 78%. All-cause mortality at 1 year was 15%; median follow-up duration was 36 (IQR 18-60) months during which 47% of patients died. One hundred and three patients had ≥1 type of BVD (44%), which predominantly consisted of NSVD (39%, mostly because of ≥moderate patient-prosthesis mismatch). BVD during follow-up included endocarditis (3.4%), clinical valve thrombosis (3.4%) and SVD (1.3%). Both endocarditis and clinically apparent valve thrombosis occurred early and late after TAVI and resulted in valve-related deaths in 38% and 13% of patients, respectively. Overall, ≥moderate haemodynamic valve deterioration occurred in 5.5% and bioprosthetic failure in 7.3%, leading to valve-related deaths in 36% of cases.
Conclusions: BVD represents a relevant health issue after TAVI with a mechanically-expandable valve. Serious but reversible causes of BVD include endocarditis and clinically apparent valve thrombosis, both carrying a time-independent hazard post-TAVI.