Introduction: The Fourth Universal Definition of Myocardial Infarction Global Taskforce recommends the use of high sensitive troponin (hs-Tn) assays in the diagnosis of acute myocardial infarction. We evaluated the analytical performance of the Atellica IM High-sensitivity Troponin I Assay (hs-TnI) (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) and compared its performance to other hs-TnI assays (Siemens Advia Centaur, Dimension Vista, Dimension EXL, and Abbott Architect (Wiesbaden, Germany)) at one or more sites across Europe.
Materials and methods: Precision, detection limit, linearity, method comparison, and interference studies were performed according to Clinical and Laboratory Standards Institute protocols. Values in 40 healthy individuals were compared to the manufacturer's cut-offs. Sample turnaround time (TAT) was examined.
Results: Imprecision repeatability CVs were 1.1-4.7% and within-lab imprecision were 1.8-7.6% (10.0-25,000 ng/L). The limit of blank (LoB), detection (LoD), and quantitation (LoQ) aligned with the manufacturer's values of 0.5 ng/L, 1.6 ng/L, and 2.5 ng/L, respectively. Passing-Bablok regression demonstrated good correlations between Atellica IM analyser with other systems; some minor deviations were observed. All results in healthy volunteers fell below the 99th percentile URL, and greater than 50% of each sex demonstrated values above the LoD. No interference was observed for biotin (≤ 1500 µg/L), but a slight bias at 5.0 g/L haemoglobin and 50 ng/L Tn was observed. TAT from was fast (mean time = 10.9 minutes) and reproducible (6%CV).
Conclusions: Real-world analytical and TAT performance of the hs-TnI assay on the Atellica IM analyser make this assay fit for routine use in clinical laboratories.
Keywords: analytical techniques and equipment; high-sensitive troponin I; immunoassay; performance evaluation.
Croatian Society of Medical Biochemistry and Laboratory Medicine.