A biosensitive analysis method development and validation was performed for accurate and rapid quantification of cefdinir (CDR) in human plasma by a liquid chromatography-tandem mass spectrometry technique coupled with electrospray ionization. Analysis was carried out using a C18 column with a flow rate of 1.0 mL/min and operating temperature of 30.0 ± 1°C. The drug was eluted by optimizing the m/z ratios of 396.20 → 227.20 and 428.17 → 241.10, for cefdinir and IS (internal standard), respectively. The intraday precision (%CV) for Cefdinir ranged from 2.8% and 6.7% as lower limit of quantification of quality control (LLOQ QC) and higher level of quantification of quality control (HQC QC), respectively, whereas these value were found to be as 3.0% and 5.6% for LLOQ and HQC, respectively after interday precision. Moreover, accuracy ranged from 107.70% (HQC QC) to 95.5% (LLOQ QC). The extraction mean recovery was found to be 83.91 ± 6.0% for cefdinir and 76.7 ± 6.23% for IS. The drug was stable throughout the analysis period. It was possible to analyze several plasma samples every day since each sample took <2.5 min to run. The method demonstrated successful quantification of CDR in human plasma, followed by pharmacokinetic profiles that were simple, accurate, sensitive and cost-effective.