Introduction: Emicizumab markedly shortens the activated partial thromboplastin time (aPTT), resulting in inaccurate measurements of procoagulant and anticoagulant factor activities. We have recently reported that mixtures of two different anti-idiotype monoclonal antibodies against emicizumab (anti-emicizumab-mAbs) allow measurement of factor (F)VIII activity (FVIII:C) and FVIII inhibitor in emicizumab-containing plasmas. It is unknown whether anti-emicizumab mAbs can work for other aPTT-based procoagulant and anticoagulant assays.
Aim: To investigate whether anti-emicizumab mAbs were measured by all of the aPTT-based assays tested.
Methods: Two anti-emicizumab-mAbs (300 μg/mL each) were preincubated with emicizumab (200 μg/mL)-spiked FVIII-deficient plasma; we then measured FVIII:C, FIX:C, FXI:C, FXII:C, protein (P)C:C, PS:C, global PC-FV (aPTT-based), and prothrombin time (PT), diluted Russel's viper venom time (dRVVT), chromogenic-based FVIII:C, FIX:C and PC:C (non-aPTT-based). Emicizumab (100 μg/mL)-spiked haemophilia (H)A plasmas from patients (n = 23) were also measured.
Results: Emicizumab shortened the clotting time in all aPTT-based assays, resulting in high levels of FVIII:C, FIX:C, FXI:C and FXII:C; low levels of PC:C and PS:C; and false-positive results for activated PC resistance. The addition of anti-emicizumab-mAbs to emicizumab-added plasma restored all factors to the initial levels without emicizumab. Chromogenic FVIII:C measurement by human FIXa/FX was affected by emicizumab, but anti-emicizumab mAbs cancelled this effect. PT-based assays and dRVVT, chromogenic FIX:C and PC:C assays showed no effect with emicizumab. Twenty-three plasma samples from HA patients also showed similar patterns.
Conclusion: Anti-emicizumab mAbs in vitro could cancel the effect of emicizumab, irrespective of the test base, resulting in accurate measurements of procoagulant and anticoagulant factor activity.
Keywords: aPTT; emicizumab; haemophilia A; idiotype antibody; monitoring.
© 2022 John Wiley & Sons Ltd.