Purpose: We report the feasibility, experience, and early outcomes of the combined intracavitary and interstitial dedicated applicator using the Kelowna GYN template (Varian, Palo Alto, CA).
Methods and materials: The Kelowna GYN template is CT compatible and used for the treatment of gynecologic cancers. In cases with patients that have an intact uterus, a modified applicator system using the Kelowna GYN template and a 3D printed adapter piece allows for compatibility with an intrautaerine tandem.
Results: We reviewed the treatment course of 23 patients comprising of 86 fractions of HDR treatment. Median D90 for cervical tumors (n = 7) was 82.4 Gy (range 77.7-92.6); for postoperative cervical tumors (n = 2) was 73.9 Gy (range 72.0-5.8); for vaginal tumors (n = 4) was 85.8 Gy (range 79.8-88.1); for recurrent endometrial (n = 10) was 86.9 Gy (range 74.8-103.2). Median EQD2 D2cc for bladder was 72.4 Gy (range 47.7-99.4), for rectum was 61.2 Gy (range 52.4-80.6), and for sigmoid colon of 50.5 Gy (44.3-66.9). At a median follow-up of 12 months, 2 patients had a local recurrence. Two patients had distant recurrence: one with carcinomatosis at 6 months, and one with pulmonary metastases at 3 months. No patients had late grade three toxicities.
Conclusions: Our single institutional experience supports the use of the Kelowna template as a robust system as a combined IC-IS applicator resulting in versatile and reproducible implants for a variety of gynecologic malignancies.
Keywords: 3d printed custom piece; Cervical cancer; Combined intracavitary and interestitial brachytherapy; Gynecologic malignancies; Interestitial brachytherapy; Kelowna template; Postoperative cervical cancer; Recurrent endometrial cancer; Vaginal cancer.
Copyright © 2022. Published by Elsevier Inc.