Objective: The main concern about measuring the concentration of rivaroxaban by anti-Xa assay in some laboratories is the lack of a commercial specific calibrator in emergencies. Therefore, this study aimed at providing a homemade anti-Xa calibrator and commercial low molecular weight heparin (LMWH) anti-Xa calibrator.
Methods: The anti-Xa plasma concentration of rivaroxaban was measured in 70 patients using a commercial specific anti-Xa calibrator, a commercial LMWH anti-Xa calibrator, and a homemade anti-Xa calibrator.
Results: We demonstrated a significant correlation and agreement (P < .001) between LMWH-calibrated anti-Xa and the commercial specific calibrator. A significant correlation (P < .001) was found between homemade calibrated anti-Xa made by normal pooled plasma and that calibrated with a commercial specific drug. The nonspecific homemade and LMWH calibrators had excellent agreement (P < .001) and can be used interchangeably.
Conclusion: Our data showed that for estimating rivaroxaban concentrations, the LMWH calibrator could be used as an alternative calibrator in the anti-Xa assay.
Keywords: Anti-Xa; NPP; homemade calibrator made by spiked normal pooled plasma; laboratory; low molecular weight heparin; rivaroxaban.
© The Author(s) 2022. Published by Oxford University Press on behalf of American Society for Clinical Pathology. All rights reserved. For permissions, please e-mail: [email protected].