One year follow-up outcomes of endovascular treatment for aortic dissection with a partial micropore stent graft in which fixation of the stent is done in zone 0: A pivotal trial first in human

Eur J Radiol. 2023 Apr:161:110720. doi: 10.1016/j.ejrad.2023.110720. Epub 2023 Feb 2.

Abstract

Objective: This study aimed to verify the efficacy and safety of the treatment for patients diagnosed with DeBakey type I and type III thoracic aortic dissections using a partial micropore stent graft.

Methods: We conducted a retrospective analysis of 32 patients who suffered from thoracic aortic dissection and underwent endovascular repair using a partial micropore stent graft at our center between December 2018 and January 2020.

Results: The technical success rate for 32 patients was 100 %, while no 30-day mortality was observed. In the 30 patients finished follow-ups, 30 (mean: 1 per patient) micropore stents were implanted, while the openings of 90 (mean: 3 per patient) aortic arch branches were covered by the stents. After more than 12 months follow-up, 26 (86.7 %) of the 30 patients presented with a complete thrombosis in the false lumen, and 4 (13.3 %) patients presented with a partial thrombosis in the false lumen. All 90 aortic arch branches were patent. No aortic arch branch artery stenosis or occlusion was observed.

Conclusions: The outcomes obtained during 12 months of follow-up suggested that performing endovascular repair for thoracic aortic dissection patients with a partial micropore stent graft is safe and effective, maintaining the patency of aortic arch branch vessels.

Keywords: Aortic arch branches; Endovascular repair; Partial micropore stent graft; Patency rate; Thoracic aortic dissection involving zone 0.

MeSH terms

  • Aorta, Thoracic
  • Aortic Aneurysm, Thoracic* / diagnostic imaging
  • Aortic Aneurysm, Thoracic* / etiology
  • Aortic Aneurysm, Thoracic* / surgery
  • Aortic Dissection* / diagnostic imaging
  • Aortic Dissection* / surgery
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation* / adverse effects
  • Endovascular Procedures*
  • Follow-Up Studies
  • Humans
  • Prosthesis Design
  • Retrospective Studies
  • Stents / adverse effects
  • Treatment Outcome