To compare the radiation dose and diagnostic ability of the 100-kVp protocol, based on the contrast noise ratio (CNR) index, during coronary artery bypass graft (CABG) vessels with those of the 120-kVp protocol. For the 120-kVp scans (150 patients), the targeted image level was set at 25 Hounsfield units (HU) (CNR120 = iodine contrast/25 HU). For the 100-kVp scans (150 patients), the targeted noise level was set at 30 HU to obtain the same CNR as in the 120-kVp scans (i.e. using 1.2-fold higher iodine contrast, CNR100 = 1.2 × iodine contrast/(1.2 × 25 HU) = CNR120). We compared the CNRs, radiation doses, detection of CABG vessels and visualisation scores of the scans acquired at 120 and 100 kVp, respectively. At the same CNR, the 100-kVp protocol may help reduce the radiation dose by ⁓30% compared with the 120-kVp protocol, without degradation of diagnostic ability during CABG.
© The Author(s) 2023. Published by Oxford University Press. All rights reserved. For Permissions, please email: [email protected].