New perspectives for the therapeutic drug monitoring of tacrolimus: Quantification in volumetric DBS based on an automated extraction and LC-MS/MS analysis

J Chromatogr B Analyt Technol Biomed Life Sci. 2023 May 15:1223:123721. doi: 10.1016/j.jchromb.2023.123721. Epub 2023 Apr 20.

Abstract

Volumetric microsampling devices have been developed for home-based capillary blood sampling and are now increasingly proposed for the therapeutic drug monitoring (TDM) of immunosuppressive drugs. Our objective was to validate a LC-MS/MS method for tacrolimus quantification based on both a manual and an automated extraction of dried blood spots (DBS) collected with a volumetric microsampling device. DBS collection was performed by placing a drop of whole blood (WB) pre-spiked with tacrolimus onto a sealing film and placing the hemaPEN® device (Trajan Scientific and Medical, Melbourne, Australia) into the drop according to the device specifications. Tacrolimus was quantified using a fully automatic preparation module connected to a LCMS system (CLAM-3020® and LCMS-8060®, Shimadzu, Marne-la-Vallée, France). The method was validated analytically and clinically in accordance with the EMA and IATDMCT guidelines. The method was linear from 1 to 100 µg/L. Within- and between-run accuracy and precision fulfilled the validation criteria (biases and imprecision <15% or 20% for the lower limit of quantification). No hematocrit effect, matrix effect or carry-over was observed. No selectivity issue was identified and dilution integrity was confirmed. Tacrolimus in DBS was stable for 14 days at room temperature and +4°C, and for 72h at +60°C. There was a good correlation between tacrolimus concentrations measured in WB and in DBS of 20 kidney and liver transplant recipients (r=0.93 and 0.87, for manual and automated extraction respectively). A method for tacrolimus measurement in DBS collected with volumetric micro-sampling device, based on a fully automated process from pre-treatment to LC-MS/MS analysis was developed and validated according to analytical and clinical criteria. This performing sampling and analytical procedure opens the perspective of an easier, faster and more efficient TDM of tacrolimus for patients, clinicians and laboratories.

Keywords: Automated extraction; LC-MS/MS; Tacrolimus; Therapeutic drug monitoring; Volumetric dried blood spots.

MeSH terms

  • Chromatography, Liquid / methods
  • Dried Blood Spot Testing / methods
  • Drug Monitoring* / methods
  • Humans
  • Immunosuppressive Agents
  • Tacrolimus*
  • Tandem Mass Spectrometry / methods

Substances

  • Tacrolimus
  • Immunosuppressive Agents