Liver T1 Relaxation Quantification Using a 3-Dimensional Interleaved Look-Locker Acquisition With T2 Preparation Pulse Sequence (3D-QALAS): Comparison With Conventional 2-Dimensional MOLLI

J Comput Assist Tomogr. 2023 May-Jun;47(3):350-354. doi: 10.1097/RCT.0000000000001409. Epub 2022 Dec 27.

Abstract

Background: Changes in liver magnetic resonance imaging T1 relaxation times are associated with histologic inflammation and fibrosis.

Objective: To compare liver T1 measurements obtained using a novel single-breath-hold 3-dimensional (3D) whole-liver T1 estimation method (3D-QALAS) to standard-of-care 2-dimensional (2D) modified Look-Locker (2D-MOLLI) measurements.

Methods: With institutional review board approval, research magnetic resonance imaging examinations were performed in 19 participants at 1.5 T. T1 relaxometry of the liver was performed using a novel 3D whole-liver T1 estimation method (3D-QALAS) as well as a 2D modified Look-Locker (2D-MOLLI) method. The 3D method covered the entire liver in a single breath hold, whereas 2D imaging was performed at 4 anatomic levels in 4 consecutive breath holds. T1 measurements from parametric maps were obtained by a single operator, and region-of-interest area-weighted mean T1 values were calculated. Pearson correlation ( r ) was used to assess correlation between T1 estimation methods, and the paired t test and Bland-Altman analysis were used to compare agreement in T1 measurements.

Results: In 18 participants (1 participant was excluded from analysis because of respiratory motion artifacts on 3D-QALAS images), 2D-MOLLI and 3D-QALAS mean T1 measurements were strongly correlated ( r = 0.95, [95% CI: 0.87-0.98]; P < 0.0001). 2D-MOLLI T1 values were significantly longer than 3D-QALAS values (647.2 ± 87.3 milliseconds vs. 554.7 ± 75.8 milliseconds; P < 0.0001) with mean bias = 92.5 milliseconds (95% limits of agreement, 36.8, 148.2 milliseconds).

Conclusion: Whole-liver T1 measurements obtained using a novel single-breath-hold 3D T1 estimation method correlate with a standard-of-care multiple consecutive-breath-hold 2D single-slice method but demonstrate systematic bias that should be considered or corrected when used in a clinical or research setting.