Purpose: For a new tibial insert design with ball-in-socket (B-in-S) medial conformity (MC), posterior cruciate ligament (PCL) retention, and flat lateral articular surface (B-in-S MC + PCL), this study determined whether internal tibial rotation and knee flexion were limited and clinical outcome scores were lower during weight-bearing activities relative to an insert with intermediate (I) (i.e., less than ball-in-socket) medial conformity (I MC + PCL).
Methods: Twenty-five patients were treated with bilateral unrestricted, caliper-verified kinematic alignment (KA) total knee arthroplasty (TKA) with an I MC + PCL insert and B-in-S MC + PCL insert in opposite knees. Each patient performed weight-bearing deep knee bend, step up, and chair rise under single-plane fluoroscopy. Analysis following 3D model-to-2D image registration determined internal tibial rotation. For each TKA, knee flexion was measured and patients completed clinical outcome scoring questionnaires.
Results: Internal tibial rotation did not differ between conformities during chair rise and step up (p = 0.3419 and 0.1030, respectively). During deep knee bend, internal tibial rotation between 90° and maximum flexion was 3° greater in the B-in-S MC + PCL group (18° vs 15°) (p = 0.0290). Mean knee flexion (p = 0.3115) and median Forgotten Joint Score (FJS), Oxford Knee Score (OKS), and Western Ontario and McMasters Universities Arthritis Index (WOMAC) scores (p = 0.2100, 0.2154, and 0.4542, respectively) did not differ between conformities.
Conclusion: An insert with ball-in-socket medial conformity, which maximizes anteroposterior (AP) stability, did not limit internal tibial rotation and knee flexion and did not lower patient-reported outcomes when implanted with unrestricted caliper-verified KA and PCL retention. The high AP stability provided by the medial ball-in-socket might interest those surgeons exploring the treatment of the active patient with a desire to return to high-level and athletic activities.
Keywords: Flat lateral articular surface; Lowest point method; Medial pivot; Single-plane fluoroscopy; Tibiofemoral kinematics; Total knee replacement.
© 2023. The Author(s) under exclusive licence to SICOT aisbl.