Side-by-side deployment of gore excluder legs at a narrow terminal aorta for endovascular aneurysm repair

J Artif Organs. 2024 Sep;27(3):238-246. doi: 10.1007/s10047-023-01406-y. Epub 2023 May 25.

Abstract

Purpose: A narrow terminal aorta is a risk factor for endograft occlusion after endovascular aneurysm repair. To minimize limb complications, we used Gore Excluder legs positioned side-by-side at the terminal aorta. We investigated the outcomes of our strategy for endovascular aneurysm repair in patients with a narrow terminal aorta.

Methods: We enrolled 61 patients who underwent endovascular aneurysm repair with a narrow terminal aorta (defined as < 18 mm in diameter) from April 2013 to October 2021. The standard procedure involves complete treatment with the Gore Excluder device. When other types of main body endografts were used, they were deployed proximal to the terminal aorta, and we used the Gore Excluder leg device in the bilateral limbs. Postoperatively, the intraluminal diameter of the legs at the terminal aorta was measured to assess the configuration.

Results: During the follow-up period (mean: 2.7 ± 2.0 years), there were no aorta-related deaths, endograft occlusions, or leg-related re-interventions. There were no significant differences between the pre- and postoperative ankle-brachial pressure index values in the dominant and non-dominant legs (p = 0.44 and p = 0.17, respectively). Postoperatively, the mean difference rate (defined as [dominant leg diameter-non-dominant leg diameter]/terminal aorta diameter) was 7.5 ± 7.1%. The difference rate was not significantly correlated with the terminal aortic diameter, calcification thickness, or circumferential calcification (r = 0.16, p = 0.22; r = 0.07, p = 0.59; and r = - 0.07, p = 0.61, respectively).

Conclusions: Side-by-side deployment of Gore Excluder legs produces acceptable outcomes for endovascular aneurysm repair with a narrow terminal aorta. The endograft expansion at the terminal aorta is tolerable without influencing calcification distribution.

Keywords: Abdominal aortic aneurysm; Endograft occlusion; Endovascular aortic repair; Gore excluder; Narrow terminal aorta.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Aneurysm, Abdominal / surgery
  • Blood Vessel Prosthesis Implantation* / adverse effects
  • Blood Vessel Prosthesis Implantation* / instrumentation
  • Blood Vessel Prosthesis Implantation* / methods
  • Blood Vessel Prosthesis*
  • Endovascular Aneurysm Repair
  • Endovascular Procedures* / adverse effects
  • Endovascular Procedures* / instrumentation
  • Endovascular Procedures* / methods
  • Female
  • Humans
  • Male
  • Middle Aged
  • Postoperative Complications
  • Prosthesis Design
  • Retrospective Studies
  • Treatment Outcome