Background: Simultaneous multi-slice diffusion-weighted imaging (SMS-DWI) can shorten acquisition time in abdominal imaging.
Purpose: To investigate the agreement and reproducibility of apparent diffusion coefficient (ADC) from abdominal SMS-DWI acquired with different vendors and different breathing schemes.
Study type: Prospective.
Subjects: Twenty volunteers and 10 patients.
Field strength/sequence: 3.0 T, SMS-DWI with a diffusion-weighted echo-planar imaging sequence.
Assessment: SMS-DWI was acquired using breath-hold and free-breathing techniques in scanners from two vendors, yielding four scans in each participant. Average ADC values were measured in the liver, pancreas, spleen, and both kidneys. Non-normalized ADC and ADCs normalized to the spleen were compared between vendors and breathing schemes.
Statistical tests: Paired t-test or Wilcoxon signed rank test; intraclass correlation coefficient (ICC); Bland-Altman method; coefficient of variation (CV) analysis; significance level: P < 0.05.
Results: Non-normalized ADCs from the four SMS-DWI scans did not differ significantly in the spleen (P = 0.262, 0.330, 0.166, 0.122), right kidney (P = 0.167, 0.538, 0.957, 0.086), and left kidney (P = 0.182, 0.281, 0.504, 0.405), but there were significant differences in the liver and pancreas. For normalized ADCs, there were no significant differences in the liver (P = 0.315, 0.915, 0.198, 0.799), spleen (P = 0.815, 0.689, 0.347, 0.423), pancreas (P = 0.165, 0.336, 0.304, 0.584), right kidney (P = 0.165, 0.336, 0.304, 0.584), and left kidney (P = 0.496, 0.304, 0.443, 0.371). Inter-reader agreements of non-normalized ADCs were good to excellent (ICCs ranged from 0.861 to 0.983), and agreement and reproducibility were good to excellent depending on anatomic location (CVs ranged from 3.55% to 13.98%). Overall CVs for abdominal ADCs from the four scans were 6.25%, 7.62%, 7.08, and 7.60%.
Data conclusion: The normalized ADCs from abdominal SMS-DWI may be comparable between different vendors and breathing schemes, showing good agreement and reproducibility. ADC changes above approximately 8% may potentially be considered as a reliable quantitative biomarker to assess disease or treatment-related changes.
Level of evidence: 2 TECHNICAL EFFICACY: Stage 2.
Keywords: abdomen; apparent diffusion coefficient; diffusion-weighted imaging; reproducibility; simultaneous multi-slice.
© 2023 International Society for Magnetic Resonance in Medicine.