Background: We analyzed real-world evidence to assess the performance of the mylife CamAPS FX hybrid closed-loop system.
Methods: Users from 15 countries across different age groups who used the system between May 9, 2022, and December 3, 2022, and who had ≥30 days of continuous glucose monitor data, and ≥30% of closed-loop usage were included in the current analysis (N = 1805).
Results: Time in range (3.9-10 mmol/L) was 72.6 ± 11.5% (mean ± SD) for all users and increased by age from 66.9 ± 11.7% for users ≤6 years old to 81.8 ± 8.7% for users ≥65 years. Time spent in hypoglycemia (<3.9 mmol/L) was 2.3% [1.3, 3.6] (median [interquartile range]). Mean glucose and glucose management indicator were 8.4 ± 1.1 mmol/L and 6.9%, respectively. Time using closed-loop was high at 94.7% [90.0, 96.9].
Conclusions: Glycemic outcomes from the present real-world evidence are comparable to results obtained from previous randomized controlled studies and confirm the efficacy of this hybrid closed-loop system in real-world settings.
Keywords: artificial pancreas; closed-loop systems; real-world evidence; type 1 diabetes.