Background: High-power, short-duration (HPSD) radiofrequency ablation (RFA) reduces procedure time; however, safety and efficacy thresholds vary with catheter design.
Objective: The study sought to determine optimal HPSD ablation conditions with a novel flexible-tipped, contact force-sensing RFA catheter.
Methods: RFA lesions were created in thigh muscle (16 swine) over a range of conditions (51-82 W, 2-40 g, 8-40 mL/min irrigation). An intracardiac study was performed (12 swine) to characterize steam pop thresholds. Lesions were created in a second intracardiac study (14 swine, n = 290 pulmonary vein isolation [PVI] lesions) with combinations of radiofrequency power, duration, and contact force. PVI was tested, animals were sacrificed, and lesions were measured.
Results: The likelihood of coagulation formation in the thigh model was <20% when power was ≤79 W, when contact force was ≤40 g, when duration was ≤11 seconds, and when irrigation rates were 8 to 40 mL/min. The impact of contact force on lesion safety and efficacy was more pronounced using HPSD (60 W/8 seconds) compared with conventional ablation (30 W/45 seconds) (P = .038). During PVI, focal atrial lesions ranged in width from 4.2 to 12.5 mm and were transmural 80.8% of the time. PVI was achieved in 13 of 14 veins. Logistic regression identified that the optimal parameters for radiofrequency application were 60 to 70 W with a duration <8 seconds and <15 g contact force.
Conclusions: Optimal HPSD lesions with this this flexible-tipped, force-sensing RFA catheter were created at 60 to 70 W for <8 seconds with <15 g contact force. Chronic studies are ongoing to assess radiofrequency parameter refinements and long-term lesion durability using these conditions.
Keywords: Atrial fibrillation; Biophysics; Catheter ablation; Contact force; Lesion formation; Radiofrequency ablation.