Background: The wearable cardioverter-defibrillator (WCD) prevents sudden cardiac death due to ventricular tachycardia (VT) or ventricular fibrillation (VF) but does not pace for post-shock asystole (PS-A) or bradycardia (PS-B;<50 beats/ min).
Objectives: The purpose of this study was to assess PS-A and PS-B in patients dying out of hospital (OOH) while wearing a WCD.
Methods: The database of the U.S. Food and Drug Administration Manufacturers and User Facility Device Experience (MAUDE) was queried for manufacturers' reports of OOH deaths while patients were wearing a WCD. Excluded were patients who did not receive a shock or were initially shocked for asystole or during resuscitation.
Results: From January 2017 to March 2022, 313 patients received an initial WCD shock for VF (n = 150), VT (n = 90), and non-VF/VT rhythms (n = 73). PS-A occurred in 204 patients (65.2%), and PS-B occurred in 111 (35.5%); 85 (41.7%) PS-A patients also had PS-B. Most PS-A patients (n = 185; 90.7%) had an initial shocked rhythm of VF or VT, but 19 patients (9.3%) were initially inappropriately shocked for atrial fibrillation/supraventricular tachycardia (n = 7) and idioventricular (n = 8) or sinus (n = 4) rhythm. PS-A occurred after the first WCD shock in 118 (63.8%) and after the first, second, or third shocks in 159 patients (85.9%). Seven patients had post-shock heart block. Eight patients had permanent pacemakers; 1 became nonfunctional after 1 shock, and 7 showed noncapture and/or asystole after 1 to 4 shocks.
Conclusions: Post-shock asystole appears to be common in patients who die OOH after being shocked by a WCD for VF or VT. PS-A also occurs after inappropriate WCD shocks for non-VF/VT rhythms. Implanted pacemakers may not prevent PS-A after a WCD shock. WCD backup pacing should be explored.
Keywords: asystole; defibrillator; sudden cardiac death; ventricular arrhythmias.
Copyright © 2023 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.