Real-world Accuracy of CGM in Inpatient Critical and Noncritical Care Settings at a Safety-Net Hospital

Diabetes Care. 2023 Oct 1;46(10):1825-1830. doi: 10.2337/dc23-0089.

Abstract

Objective: We sought to determine real-world accuracy of inpatient continuous glucose monitoring (CGM) at multiple levels of acuity in a large safety-net hospital.

Research design and methods: We analyzed records from hospitalized patients on Dexcom G6 CGM, including clinical, point of care (POC), and laboratory (Lab) glucose, and CGM data. POC/Lab values were matched to the closest timed CGM value. Encounters were divided into not critically ill (NCI) versus critically ill (CI). CGM accuracy was evaluated.

Results: Paired readings (2,744 POC-CGM; 3,705 Lab-CGM) were analyzed for 233 patients with 239 encounters (83 NCI, 156 CI). POC-CGM aggregated and average mean absolute relative differences (MARD) were 15.1% and 17.1%. Lab-CGM aggregated and average MARDs were 11.4% and 12.2%. Accuracy for POC-CGM and Lab-CGM was 96.5% and 99.1% in Clarke Error Grid zones A/B.

Conclusions: Real-world accuracy of inpatient CGM is acceptable for NCI and CI patients. Further exploration of conditions associated with lower CGM accuracy in real-world settings is warranted.