Introduction: We studied the impact of the use of three-dimensional multidetector computed tomography (3D-MDCT) and fluoroscopy fusion on percutaneous left atrial appendage occlusion (LAAO) procedures in relation to procedure time, contrast volume, fluoroscopy time, and total radiation.
Methods: This was a single-center, prospective, single-blinded, randomized control trial. Patients meeting criteria for LAAO were randomized to undergo LAAO with the WATCHMAN FLXTM device with and without 3D-MDCT-fluoroscopy fusion guidance using a prespecified protocol using computed tomography angiography for WATCHMAN FLXTM sizing, moderate sedation, and intracardiac echocardiography for procedural guidance.
Results: Overall, 59 participants were randomly assigned to the fusion (n = 33) or no fusion (n = 26) groups. The median (interquartile range) age was 79 (75-83) years, 24 (41%) were female, and 55 (93%) were Caucasian. The median CHA2 DS2 VASc and HASBLED scores were 5 (4-6) and 3 (3-4), respectively. At the time of the study, 51 (53%) patients were on a direct acting oral anticoagulant. There were no significant differences between the fusion and no fusion groups in procedure time (52.4 ± 15.4 vs. 56.8 ± 19.5 min, p = .36), mean contrast volume used (33.8 ± 12.0 vs. 29.6 ± 11.5 mls, p = .19), mean fluoroscopy time (31.3 ± 9.9 vs. 28.9 ± 8.7 min, p = .32), mean radiation dose (1177 ± 969 vs. 1091 ± 692 mGy, p = .70), and radiation dose product curve (23.9 ± 20.5 vs. 35.0 ± 49.1 Gy cm2 , p = .29). There was no periprosthetic leak in the two groups in the immediate 1-month postprocedure follow-up periods.
Conclusions: There was no significant difference with and without 3D-MDCT-fluoroscopy fusion in procedure time, contrast volume use, radiation dose, and radiation dose product.
Keywords: atrial fibrillation; computed tomography; fluoroscopy; fusion; left atrial appendage occlusion.
© 2023 Wiley Periodicals LLC.