Clinical outcomes using a 3D printed tandem-needle-template and the EMBRACE-II planning aims for image guided adaptive brachytherapy in locally advanced cervical cancer

Acta Oncol. 2023 Nov;62(11):1470-1478. doi: 10.1080/0284186X.2023.2246642. Epub 2023 Aug 18.

Abstract

Background: Extensive local disease or narrow vagina may compromise brachytherapy (BT) in patients with cervical cancer. This is the first study to analyze long-term outcomes of using 3D printed vaginal tandem-needle templates (3DP TNT) for transvaginal insertion of needles in parallel (P) or parallel and oblique (P&O) direction to the tandem.

Material and methods: All patients treated with BT using 3DP TNT from 2015-2020 were included. Decision to use a 3DP TNT and preplanning were made after 4-5 weeks of external beam radiotherapy, based on gynecological examination and MRI with a tandem-ring applicator in situ. The TNT was 3D-printed in house consisting of a circular template with P&O holes for guidance of plastic needles and a shaft fitting the uterine tandem. Thus, the radioactive source was never in direct contact with the 3DP TNT. The TNT was 3D printed in a standard or personalized configuration. Planning aims were based on the Embrace II protocol.

Results: 101 patients (median age of 63 years) were included: 49 with P needles only and 52 with P&O needles. Personalized TNT was used in 19 patients in the P&O group. Performance status (WHO) was > 0 in 48%. FIGO2018 stage III-IV was present in 77%. T-score at diagnosis and BT was 9.1 and 6.3 respectively, with a significantly higher T-score in the P&O compared to P group. The mean high-risk CTV D90 was 93 Gy with no significant difference between the two groups. Three-year local control rates were 85%, 95%, 75% for the overall, P- and P&O group respectively and 68%, 80% and 56% for cancer specific survival. Grade ≥3 treatment related complications were observed in 10 (10%) patients.

Conclusions: 3DP TNT for BT in cervical cancer provides successful management of very extensive local disease and/or unfavorable anatomy with the possibility for treatment individualization.

Keywords: 3d-printing; Cervical cancer; image guided adaptive brachytherapy; interstitial; treatment individualization.

MeSH terms

  • Brachytherapy* / methods
  • Female
  • Humans
  • Middle Aged
  • Pelvis
  • Printing, Three-Dimensional
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods
  • Uterine Cervical Neoplasms* / radiotherapy