Repeatability of pupil diameter measurements using three different topography devices

PLoS One. 2023 Aug 18;18(8):e0290417. doi: 10.1371/journal.pone.0290417. eCollection 2023.

Abstract

Purpose: To evaluate the intra- and inter-device repeatability of pupil diameter measurements using three different devices in patients prior to corneal refractive surgery.

Methods: We examined preoperative measurements from a total of 204 eyes (102 patients) scheduled for corneal refractive surgery at two private centers between July and December 2021. Three consecutive scans were performed with three different devices (Sirius anterior segment analyzer, Pentacam HR, IOLMaster 500) in the same session by the same examiner under standardized conditions. To assess the intra- and inter-device repeatability, we calculated the Intraclass Correlation Coefficient (ICC) and demonstrated results using Bland-Altman plots.

Results: The measurement accuracy (intra-device repeatability) of Sirius and IOLMaster was comparable (ICC = 0.64 and 0.61, respectively), with almost no statistically significant differences. Sirius showed the highest measurement accuracy among the three devices. Pentacam measurements resulted in lower precision, with an ICC of 0.09. The agreement between the pairs of devices (inter-device repeatability) was low (wide LoA ranges, Table 5).

Conclusion: In this study, the intra-device repeatability of Sirius and IOLMaster was higher than that of the Pentacam, although it did not achieve an optimal level across all three devices. The three devices examined cannot be used interchangeably.

MeSH terms

  • Cornea
  • Humans
  • Interior Design and Furnishings
  • Ophthalmology*
  • Pupil*

Grants and funding

The authors received no specific funding for this work.