FDA-cleared artificial intelligence and machine learning-based medical devices and their 510(k) predicate networks

Lancet Digit Health. 2023 Sep;5(9):e618-e626. doi: 10.1016/S2589-7500(23)00126-7.

Abstract

The US Food and Drug Administration is clearing an increasing number of artificial intelligence and machine learning (AI/ML)-based medical devices through the 510(k) pathway. This pathway allows clearance if the device is substantially equivalent to a former cleared device (ie, predicate). We analysed the predicate networks of cleared AI/ML-based medical devices (cleared between 2019 and 2021), their underlying tasks, and recalls. More than a third of cleared AI/ML-based medical devices originated from non-AI/ML-based medical devices in the first generation. Devices with the longest time since the last predicate device with an AI/ML component were haematology (2001), radiology (2001), and cardiovascular devices (2008). Especially for devices in radiology, the AI/ML tasks changed frequently along the device's predicate network, raising safety concerns. To date, only a few recalls might have affected the AI/ML components. To improve patient care, a stronger focus should be placed on the distinctive characteristics of AI/ML when defining substantial equivalence between a new AI/ML-based medical device and predicate devices.

Publication types

  • Review
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artificial Intelligence*
  • Humans
  • Machine Learning
  • Radiology*
  • United States
  • United States Food and Drug Administration