Recombinant monoclonal antibodies (mAbs) have been spurring the rapid growth of commercial biotherapeutics. During production their charge heterogeneity must be assessed as a critical quality attribute to ensure safety, efficacy, and potency. Although imaged capillary isoelectric focusing (icIEF) is a powerful tool for this process, it could be improved further with tandem high-resolution mass spectrometry (HRMS). In this work, a nano-electrospray ionization (nano-ESI) apparatus was constructed to directly couple icIEF to HRMS. The system was evaluated with the standard NISTmAb, as well as more complex mAb, bi-specific antibody, and fusion protein samples. NISTmAb concentrations as low as 0.25 mg/ml demonstrated excellent sensitivity. There were good repeatabilities at 1 mg/ml with 7.58% and 8.01% RSDs for intention time and MS intensity, respectively, and the HRMS signal showed a strong linearity (R = 0.9983) across different concentrations. Meanwhile, the fingerprinting of the complex samples illustrated the versatility and potential of icIEF-HRMS. icIEF-HRMS developed can provide a comprehensive understanding of the underlying structural modifications that impact protein charge heterogeneity. Compared to the traditional ESI, nano-ESI can significantly improve sensitivity while maintaining a reasonable repeatability and throughput. Furthermore, the interface is much easier to connect, and is compatible with many commercial HRMS instruments.
Keywords: Bi-specific antibody; Fusion protein; Imaged capillary isoelectric focusing; Low-flow electrospray ionization; Mass spectrometry; Monoclonal antibodies; Nano electrospray ionization; Protein heterogeneity.