A minimally-invasive manipulator characterized by hyper-redundant kinematics and embedded sensing modules is presented in this work. The bending angles (tilt and pan) of the robot tip are controlled through tendon-driven actuation; the transmission of the actuation forces to the tip is based on a Bowden-cable solution integrating some channels for optical fibers. The viability of the real-time measurement of the feedback control variables, through optoelectronic acquisition, is evaluated for automated bending of the flexible endoscope and trajectory tracking of the tip angles. Indeed, unlike conventional catheters and cannulae adopted in neurosurgery, the proposed robot can extend the actuation and control of snake-like kinematic chains with embedded sensing solutions, enabling real-time measurement, robust and accurate control of curvature, and tip bending of continuum robots for the manipulation of cannulae and microsurgical instruments in neurosurgical procedures. A prototype of the manipulator with a length of 43 mm and a diameter of 5.5 mm has been realized via 3D printing. Moreover, a multiple regression model has been estimated through a novel experimental setup to predict the tip angles from measured outputs of the optoelectronic modules. The sensing and control performance has also been evaluated during tasks involving tip rotations.
Keywords: hyper-redundant manipulator; medical robotics; multiple regression model; optoelectronic sensor.