Objective: Advanced hybrid closed-loop systems (AHCL) have been shown to improve glycemic control and patient-reported outcomes in type 1 diabetes. The aim was to analyze the outcomes of two commercially available AHCL in real life.
Research design and methods: A prospective study was performed, including adolescents and adults with type 1 diabetes, AHCL naïve, from 14 centers, who initiated the use of MM780G with SmartGuard or Tandem t:slimX2 with Control-IQ. Baseline and 3-month evaluations were performed, assessing HbA1c, time in different glycemic ranges, and patient-reported outcomes. The primary outcome was the between-group time in range 70-180 mg/dL difference from beginning to end of follow-up.
Results: One hundred fifty participants were included, with 75 initiating each system (age: 39.9 ± 11.4 years [16-72]; 64% female; diabetes duration: 21.6 ± 11.9 years). Time in range increased from 61.53 ± 14.01% to 76.17 ± 9.48% (P < 0.001), with no between-group differences (P = 0.591). HbA1c decreased by 0.56% (95% CI 0.44%, 0.68%) (6 mmol/mol, 95% CI 5, 7) (P < 0.001), from 7.43 ± 1.07% to 6.88 ± 0.60% (58 ± 12 to 52 ± 7 mmol/mol) in the MM780G group, and from 7.14 ± 0.70% to 6.56 ± 0.53% (55 ± 8 to 48 ± 6 mmol/mol) in the Control-IQ group (both P < 0.001 to baseline, P = 0.819 between groups). No superiority of one AHCL over the other regarding fear of hypoglycemia or quality of life was found. Improvement in diabetes-related distress was higher in Control-IQ users (P = 0.012). Sleep quality was improved (PSQI: from 6.94 ± 4.06 to 6.06 ± 4.05, P = 0.004), without differences between systems. Experience with AHCL, evaluated by the INSPIRE measures, exceeded the expectations.
Conclusions: The two AHCL provide significant improvement in glucose control and satisfaction, with no superiority of one AHCL over the other.
© 2024 by the American Diabetes Association.