Background: Transcatheter aortic valve implantation is an effective treatment for patients with aortic stenosis; however, complications related to paravalvular leakage (PVL) persist, including increased risk of mortality, cardiovascular mortality, and rehospitalisation.

Aims: We sought to evaluate the clinical outcomes and valve performance at 1 year in patients with severe aortic stenosis treated with the ACURATE neo2 valve in a post-market clinical setting.

Methods: Valve Academic Research Consortium-2 safety events were assessed up to 1 year. Independent core laboratories evaluated echocardiographic measures of valve performance and hypoattenuated leaflet thickening (HALT; as measured by four-dimensional computed tomography).

Results: The study enrolled 250 patients (64% female; mean age: 81 years; baseline Society of Thoracic Surgeons risk score: 2.9±2.0%); 246 patients were implanted with ACURATE neo2. All-cause mortality was 0.8% at 30 days and 5.1% at 1 year. The 1-year rates for stroke and disabling stroke were 3.0% and 1.3%, respectively. Overall, HALT of >50% leaflet involvement of at least one leaflet was present in 9% of patients at 30 days and in 12% of patients at 1 year. No association was observed between the presence of HALT and 1-year clinical or haemodynamic outcomes. Early haemodynamic improvements were maintained up to 1 year (mean aortic valve gradient: 47.6±14.5 mmHg at baseline, 7.6±3.2 mmHg at 1 year; mean aortic valve area: 0.7±0.2 cm2 at baseline, 1.7±0.4 cm2 at 1 year). At 1 year, 99% of patients had mild or no/trace PVL (<1% had moderate PVL; no patient had severe PVL).

Conclusions: The study outcomes confirm favourable performance and safety up to 1 year in patients treated with ACURATE neo2 in routine clinical practice. (ClinicalTrials.gov: NCT04655248).

Publication types

  • Clinical Study

MeSH terms

  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / surgery
  • Female
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Prosthesis Design
  • Stroke* / etiology
  • Transcatheter Aortic Valve Replacement* / methods
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT04655248