Ongoing Geometric Remodeling of the Parent Artery After Flow-Diverter Stent Reconstruction in Cerebral Aneurysms: The Device Design Matters

World Neurosurg. 2024 Feb:182:e597-e601. doi: 10.1016/j.wneu.2023.11.153. Epub 2023 Dec 3.

Abstract

Objective: Configuration changes of the parent artery (PA) after flow-diverter (FD) stent reconstruction, caused by the bending force of the device, may have an additional role in aneurysm occlusion as a result of the secondary alteration of intra-aneurysmal hemodynamics related to the geometry alteration of the vessel. To determine the degree of PA deformation and aneurysm occlusion rates after deployment of 2 different types of FD.

Methods: Patients treated with 2 different designs of cobalt-chromium braid (48 and 64 wire braid) structure FD were subject to analysis. Vascular angle changes at the level of the reconstructed segment immediately after FD deployment and at 1 year follow-up were measured and the potential relationship with aneurysmal occlusion rate was analyzed.

Results: Forty-two patients harboring 48 aneurysms were included in the present study. The aneurysms were divided into side wall (85.4%) and bifurcation types (14.6%). Twenty-six aneurysms were treated using the Pipeline FD (48 wire braid; 54.2%) and 22 using the Evolve FD (64 wire braid; 45.8%). Of the 48 aneurysms, 42 (87.5%) met the primary end point of complete occlusion at 12 months. The median postdeployment angle change was 7.04°± 4.59° for the Pipeline and 5.05°± 2.49° for the Evolve, whereas the median 12 months follow-up angle change was 15.49°± 10.99° and 10.01°± 8.83°, respectively. PA angle changes were significantly higher in the bifurcation group compared with the side wall group both during procedure and at 12 months follow-up. Angle change had a statistically nonsignificant association with complete aneurysm occlusion.

Conclusions: PA deformation starts immediately after deployment and remodeling continues for 1 year after. Aneurysms located in the vessel bifurcation were more prone to PA straightening after FD deployment than were side wall aneurysms. Furthermore, Pipeline seemed to be more prone to inducing vascular deformation, compared with Evolve.

Keywords: Flow-diverter devices; Intracranial aneurysms; Parent artery deformation.

MeSH terms

  • Arteries
  • Embolization, Therapeutic* / methods
  • Endovascular Procedures* / methods
  • Equipment Design
  • Humans
  • Intracranial Aneurysm* / diagnostic imaging
  • Intracranial Aneurysm* / etiology
  • Intracranial Aneurysm* / surgery
  • Stents
  • Treatment Outcome