The study investigated radiation dose, vascular computed tomography (CT) enhancement and image quality of cardiac computed tomography angiography (CCTA) with and without bolus tracking (BT) methods in infants with congenital heart disease (CHD). The volume CT dose index (CTDIvol) and dose length product (DLP) were recorded for all CT scans, and the effective dose was obtained using a conversion factors. The CT number for the ascending aorta (AO) and pulmonary artery (PA), image noise of muscle tissue and contrast-to-noise ratio (CNR) were measured and calculated. The median values in the groups with and without BT were 2.20 mGy versus 0.44 mGy for CTDIvol, 8.10 mGy·cm versus 6.20 mGy·cm for DLP, and 0.66 mSv versus 0.51 mSv for effective dose (p < 0.001). There were no statistical differences in vascular CT enhancement, image noise, and CNR. CCTA without BT methods can reduce the radiation dose while maintaining vascular CT enhancement and image quality compared to CCTA with BT methods.
© The Author(s) 2023. Published by Oxford University Press. All rights reserved. For Permissions, please email: [email protected].