Comparing a 19-gauge fine-needle biopsy needle with a 22-gauge fine-needle biopsy needle for EUS-guided liver biopsy sampling: a prospective randomized study

Gastrointest Endosc. 2024 Jun;99(6):931-937. doi: 10.1016/j.gie.2023.12.022. Epub 2023 Dec 21.

Abstract

Background and aims: EUS-guided liver biopsy (EUS-LB) sampling is being increasingly used. We performed a prospective randomized trial to compare specimen adequacy of a 19-gauge fine-needle biopsy (FNB) needle with a 22-gauge FNB Franseen tip needle for EUS-LB sampling.

Methods: Forty-two consecutive patients referred for EUS-LB sampling were prospectively randomized to a 19-gauge or 22-gauge FNB needle. When the specimen with the 22-gauge needle was macroscopically inadequate, an additional pass with the 19-gauge needle was done. Bilobar EUS-LB sampling was performed with heparinized wet suction using 1 pass and 3 actuations per lobe. Descriptive statistics were computed for all variables.

Results: Biopsy sampling was performed for abnormal liver enzymes in 95.5% of patients (57% women; average age, 51 years). Five patients undergoing sampling with the 22-gauge FNB needle had macroscopically inadequate specimens and required additional biopsy sampling with the 19-gauge FNB needle. Mean preprocessing length of the longest tissue core was 21.5 ± 6.3 mm with a 19-gauge FNB needle compared with 9.4 ± 5.5 mm with the 22-gauge FNB needle (P < .001). Postprocessing specimens were significantly longer with 19-gauge than with 22-gauge FNB needles (17.4 mm vs 6.8, P < .001). There were no adverse events, and postprocedure pain and discomfort was similar in both groups (14% for 19-gauge vs 10% for 22-gauge, P = .99).

Conclusions: Liver core biopsy sampling using the 19-gauge FNB needle is superior to the 22-gauge FNB needle in terms of length of longest core and aggregate specimen length. Considerably more fragmentation of the 22-gauge cores occurs during tissue processing. No increased postprocedure pain or AEs were found with the 19-gauge needle. A 19-gauge FNB needle is preferred to the 22-gauge FNB needle for EUS-LB. (Clinical trial registration number: NCT04806607.).

Publication types

  • Randomized Controlled Trial
  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Biopsy, Large-Core Needle / instrumentation
  • Biopsy, Large-Core Needle / methods
  • Endoscopic Ultrasound-Guided Fine Needle Aspiration* / instrumentation
  • Endoscopic Ultrasound-Guided Fine Needle Aspiration* / methods
  • Female
  • Humans
  • Liver Diseases / pathology
  • Liver Neoplasms / pathology
  • Liver* / pathology
  • Male
  • Middle Aged
  • Needles*
  • Prospective Studies

Associated data

  • ClinicalTrials.gov/NCT04806607