Evaluation of Inlet Temperature with Three Sprayer Designs for Desorption Electrospray Ionization Mass Spectrometry Tissue Analysis

J Am Soc Mass Spectrom. 2024 Feb 7;35(2):224-233. doi: 10.1021/jasms.3c00332. Epub 2024 Jan 5.

Abstract

Mass spectrometry imaging (MSI) allows for the spatially resolved detection of endogenous and exogenous molecules and atoms in biological samples, typically prepared as thin tissue sections. Desorption electrospray ionization (DESI) is one of the most commonly utilized MSI modalities in preclinical research. DESI ion source technology is still rapidly evolving, with new sprayer designs and heated inlet capillaries having recently been incorporated in commercially available systems. In this study, three iterations of DESI sprayer designs are evaluated: (1) the first, and until recently only, commercially available Waters sprayer; (2) a developmental desorption electro-flow focusing ionization (DEFFI)-type sprayer; and (3) a prototype of the newly released Waters commercial sprayer. A heated inlet capillary is also employed, allowing for controlled inlet temperatures up to 500 °C. These three sprayers are evaluated by comparative tissue imaging analyses of murine testes across this temperature range. Single ion intensity versus temperature trends are evaluated as exemplar cases for putatively identified species of interest, such as lactate and glutamine. A range of trends are observed, where intensities follow either increasing, decreasing, bell-shaped, or other trends with temperature. Data for all sprayers show approximately similar trends for the ions studied, with the commercial prototype sprayer (sprayer version 3) matching or outperforming the other sprayers for the ions investigated. Finally, the mass spectra acquired using sprayer version 3 are evaluated by uniform manifold approximation and projection (UMAP) and k-means clustering. This approach is shown to provide valuable insight that is complementary to the presented univariate evaluation for reviewing the parameter space in this study. Full spectral temperature optimization data are provided as supporting data to enable other researchers to design experiments that are optimal for specific ions.

MeSH terms

  • Animals
  • Bays*
  • Hot Temperature
  • Ions
  • Mice
  • Spectrometry, Mass, Electrospray Ionization* / methods
  • Temperature

Substances

  • Ions