Cochlear implants can provide an advanced treatment option to restore hearing. In standard pre-implant procedures, many factors are already considered, but it seems that not all underlying factors have been identified yet. One reason is the low quality of the conventional computed tomography images taken before implantation, making it difficult to assess these parameters. A novel method is presented that uses the Pietsch Model, a well-established model of the human cochlea, as well as landmark-based registration to address these challenges. Different landmark numbers and placements are investigated by visually comparing the mean error per landmark and the registrations' results. The landmarks on the first cochlear turn and the apex are difficult to discern on a low-resolution CT scan. It was possible to achieve a mean error markedly smaller than the image resolution while achieving a good visual fit on a cochlear segment and directly in the conventional computed tomography image. The employed cochlear model adjusts image resolution problems, while the effort of setting landmarks is markedly less than the segmentation of the whole cochlea. As a next step, the specific parameters of the patient could be extracted from the adapted model, which enables a more personalized implantation with a presumably better outcome.
© 2024. The Author(s).