Purpose: To report 36-month outcomes and subgroup analysis of the ABRE study evaluating the safety and effectiveness of the Abre venous self-expanding stent system for the treatment of symptomatic iliofemoral venous outflow obstruction disease.
Methods: The ABRE study was a prospective, multicenter, nonrandomized study that enrolled and implanted Abre venous stents in 200 participants (mean age 51.5 years [SD ± 15.9], 66.5% women) with symptomatic iliofemoral venous outflow obstruction at 24 global sites. Outcomes assessed through 36 months included patency, major adverse events, stent migration, stent fracture, and quality-of-life changes. Adverse events and imaging studies were adjudicated by independent clinical events committee and core laboratories, respectively.
Results: Primary, primary-assisted, and secondary patency through 36 months by Kaplan-Meier estimates were 81.6%, 84.8%, and 86.3%, respectively. The cumulative incidence of major adverse events through 36 months was 10.2%, mainly driven by 12 thrombosis events. Subgroup analyses demonstrated a primary patency of 76.5% in the acute deep vein thrombosis group, 70.4% in the postthrombotic syndrome group, and 97.1% in the nonthrombotic iliac vein lesion group through 36 months. The overall mean lesion length was 112.4 mm (SD ± 66.1). There were no stent fractures or migrations in this study. Quality of life and venous functional assessments demonstrated significant improvements from baseline to 36 months across all patient subsets.
Conclusions: Results from the ABRE study demonstrated sustained patency with a good safety profile after implantation of a dedicated venous stent in patients with symptomatic iliofemoral venous outflow obstruction disease.
Copyright © 2024 SIR. Published by Elsevier Inc. All rights reserved.