Mind the Gap in TAVR: Recognizing and Managing Misloaded Self-Expanding TAVR Devices

JACC Case Rep. 2023 Dec 29;29(3):102192. doi: 10.1016/j.jaccas.2023.102192. eCollection 2024 Feb 7.

Abstract

Misloading during transcatheter aortic valve replacement (TAVR) is rare but can cause unpredictable valve release if unrecognized. We describe how to identify a misloaded ACURATE neo2 device, and 3 methods to solve this by using a modified technique of valve deployment, ipsilateral extraction, and contralateral valve externalization with extracorporeal valve release.

Keywords: ACURATE; TAVI; misloading; self-expanding valve.

Publication types

  • Case Reports