Objectives: Patent hemostasis (PH) is essential for preventing radial artery occlusion (RAO) after trans-radial procedures; however, it remains unclear how it should be obtained. The aim of this multicenter randomized study was to evaluate whether the use of an adjustable device (AD), inflated with a pre-determined amount of air (AoA), was more effective than a non-adjustable device (non-AD) for achieving PH, thereby reducing the incidence of RAO.
Methods: We enrolled a total of 480 patients undergoing transradial procedure at 3 Italian institutions. Before the procedure, a modified Reverse Barbeau Test (mRBT) was performed in all patients to evaluate the AoA to be eventually inflated in the AD. After the procedure, patients were randomized into 2 groups: (1) AD Group, using TR-Band (Terumo) inflated with the pre-determined AoA; and 2) non-AD Group, using RadiStop (Abbott). An RBT was performed during compression to demonstrate the achievement of PH, as well as 24 hours later to evaluate the occurrence of RAO.
Results: PH was more often obtained in the AD Group compared with the non-AD Group (90% vs 64%, respectively, P less than .001), with no difference in terms of bleedings. RAO occurred more often in the non-AD Group compared with the AD Group (10% vs 3%, respectively, P less than .001). Of note, mRBT was effective at guiding AD inflation and identifying high-risk patients in whom PH was more difficult to obtain.
Conclusions: The use of AD, filled with a predetermined AoA, allowed PH significantly more often compared with non-AD, providing a significantly reduced incidence of RAO.
Keywords: patent hemostasis; percutaneous coronary intervention; radial artery occlusion; transradial percutaneous intervention.